Home » Moisture & Humidity » Moisture Analyzers » Moisture Image Series 1
Print

Moisture Image Series 1 - Moisture Analyzer

main-mis1-display

The Moisture Image Series 1 (MIS 1) analyzer is the top-of-the-line model in GE Panametrics’ family of Moisture Series analyzers. This multifunction analyzer uses charts, graphs and windows to display any of six parameters (moisture, oxygen, temperature, pressure and two auxiliary inputs) for as many as six channels simultaneously on a large, backlit LCD. Real-time data can also be exported to a PC using the instrument’s RS232 port and standard or custom software packages.

Combining the MIS 1 analyzer, the MIS probe, and our nondepleting oxygen cell creates a system capable of measuring as many as six points each of ambient to trace amounts of moisture and oxygen. Computer-enhanced response and a true ppbv scale are included for fast response and maximum resolution and accuracy in extremely low-level measurement applications.

-->For added flexibility, the analyzer’s auxiliary inputs accept measurements from any sensor with current or voltage output, making the MIS 1 a true multifunction analyzer. Other features include intrinsic safety for all inputs, isolated analog outputs, alarm relays and a built-in 24-VDC power supply for current loops. The instrument is available in rack, bench, panel, weatherproof and explosion-proof packages.


Features

  • Absolute humidity in liquids and gases
  • Ambient to ppb moisture and oxygen ranges
  • One to six measurement channels
  • Simultaneous graphic display of six measurements
  • 16-bit moisture measurement resolution
  • Compatible with MISP,TF and M Series type moisture probes
  • Auxiliary inputs for other parameters
  • Computer-enhanced speed of response
  • Intrinsically safe, certified to EEx ia IIC and Class I, Div. 1, All Groups

 

Applications


This Panametrics aluminum oxide moisture sensor probe measures moisture concentration in gases and non-aqueous liquids from trace to ambient levels. It is designed to be used in conjunction with all Panametrics moisture analyzers for industries including:
• Petrochemical
• Natural gas
• Industrial gas
• Semiconductor
• Furnace gas/heat treating
• Power generation
• Air dryer
• Pharmaceutical
• Aerospace

Specifications

Moisture Probe

Intrinsic Safety Intrinsically safe when connected to a Panametrics Moisture Series analyzer, PM880 portable hygrometer or intrinsically safe barriers in accordance with the user’s manual.
M Series moisture probe:
  • BAS01ATEX1096X
  • II 1 G EEx ia IIC T4 (-20°C to +80°C), and
  • CSA C US Class I, Division 1, Groups A,B,C&D T4,
  • LR44204-23
Optional Electrical Heater Complies with EMC Directive 89/336/EEC and PED 97/23/EC for DN<25
Type Aluminum oxide moisture sensor
Calibration Each probe is individually computer calibrated against known moisture concentrations, traceable to NIST or NPL.
Dew/Frost Point Calibration Ranges • Overall capability: 140°F to -166°F (60°C to –110°C) by request
• Standard calibration range: 68°F to -112°F (20°C to –80°C) with data to -166°F (-110°C)
• Ultralow calibration range: -58°F to -148°F (-50°C to –100°C) with data to -166°F (-110°C)
Accuracy • ±3.6°F (±2°C) in range of 140°F to -85°F (60°C to –65°C)
• ±5.4°F (±3°C) in range of -86°F to -166°F (-66°C to 110°C)
Repeatability • ±0.9°F (±0.5°C) in range of 140°F to -85°F (60°C to –65°C)
• ±1.8°F (±1.0°C) in range of -86°F to -166°F (–66°C to –110°C)
Temperature • Operating: -166°F to 158°F (–110°C to 70°C)
• Storage: Maximum 158°F (70°C)
Operating Pressure (Dependent on Mount) • Flange process connection (M1): 5 :Hg to 75 psig (6 bar)
• Threaded process connection (M2): 5 :Hg to 5000 psig (345 bar)
Flow Range • Gases: Static to 10,000 cm/s linear velocity at 1 atm
• Liquids: Static to 10 cm/s linear velocity at density of 1 g/cc
Response Time Less than five seconds for 63% step change in moisture content in either wet-up or dry-down cycle
Input Voltage 1 VAC, 77 Hz
Impedance Range 50 kS to 2 MS depending on water vapor
Limited Warranty • Calibration: Six months from delivery
• Materials and workmanship: One year from delivery

Optional Temperature Sensor

Type Nonlinear negative temperature coefficient (NTC) thermistor (resultant temperature linearized by microprocessor)
Operating Range –22°F to 158°F (–30°C to 70°C)
Accuracy ±0.9°F (±0.5°C) overall
Response Time (Maximum) 1 second in well-stirred oil or 10 seconds in still air for a 63% step change in increasing or decreasing
temperature
Optional Electrical Heater 120 VAC, 120 W
230 VAC, 75 W

Download

Descriptions English Italiano Português
BASEEFA System Drawing for MIS 1 IS System acrobat-reader
CSA-NRTL System Drawing for MIS 1 IS System acrobat-reader
M Series Moisture Probe acrobat-reader
MIS 1 Bench-Mount Package Drawing acrobat-reader
MIS 1 Interconnection Drawing for Rack, Bench & Panel acrobat-reader
MIS 1 Panel-Mount Package Drawing acrobat-reader
MIS 1 Rack-Mount Package Drawing acrobat-reader
MIS 1 Weatherproof Package Drawing acrobat-reader
MIS1 (w/703-1250 Controller Board) acrobat-reader
MIS1 (w/703-1250 Controller Board) Service Operating Manual acrobat-reader
MIS1 Hygrometer Operating Manual acrobat-reader
MIS1 Quick Start acrobat-reader
Moisture Image Series 1 Analisador de Umidade da Panametrics acrobat-reader
Moisture Image Series 1 Analizzatore di umidita Panametrics acrobat-reader
Moisture Image Series 1 Brochure acrobat-reader
Moisture Image Series moisture probe with aluminum oxide Panametrics acrobat-reader
Moisture Image Series Probe acrobat-reader
Nondepleting Oxygen Cell acrobat-reader
Three-Function Moisture Probe acrobat-reader

Click here MIS-1 Updates to download this software

acrobat-reader

 

Dew/Frost Point Calibration Ranges • Overall capability: 140°F to -166°F (60°C to –110°C) by request
• Standard calibration range: 68°F to -112°F (20°C to –80°C) with data to -166°F (-110°C)
• Ultralow calibration range: -58°F to -148°F (-50°C to –100°C) with data to -166°F (-110°C)